Abstract

We performed low-flow sevoflurane anaesthesia at a flow rate of 1 litre min-1 in three groups (n = 8 each) using 600 g of fresh soda lime (control group), 600 g of soda lime with 60 ml of water added (water group) or 600 g of soda lime carbon dioxide, that is partly lime (carbon dioxide group). Degradation products in the system were measured hourly. Inspired and end-tidal carbon dioxide and sevoflurane concentrations, carbon dioxide and temperature of the soda lime were monitored. CF2 = C(CF3)-O-CH2F (compound A) was the only sevoflurane degradation product detected. The mean maximum concentration of compound A was significantly higher in the control group (mean 16.0 (SD 5.0) ppm) than in the water (1.4 (1.0) ppm) or carbon dioxide (4.0 (1.8) ppm) group, and the maximum temperature of the soda lime was significantly lower in the carbon dioxide group (30.7 (3.5)°C) than in the control (43.4 (1.8)°C) or water (40.8 (1.8)°C) group (P < 0.05). The use of partly exhausted soda lime or soda lime with water added reduced compound A concentrations in the system during low-flow sevoflurane anaesthesia.

Cite

CITATION STYLE

APA

Moriwaki, G., Bito, H., & Ikeda, K. (1997). Partly exhausted soda lime or soda lime with water added, inhibits the increase in compound A concentration in the circle system during low-flow sevoflurane anaesthesia. British Journal of Anaesthesia, 79(6), 782–786. https://doi.org/10.1093/bja/79.6.782

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free