Changes in stroke volume induced by passive leg raising in spontaneously breathing patients: comparison between echocardiography and Vigileo/FloTrac device.

125Citations
Citations of this article
156Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

INTRODUCTION: Passive leg raising (PLR) is a simple reversible maneuver that mimics rapid fluid loading and increases cardiac preload. The effects of this endogenous volume expansion on stroke volume enable the testing of fluid responsiveness with accuracy in spontaneously breathing patients. However, this maneuver requires the determination of stroke volume with a fast-response device, because the hemodynamic changes may be transient. The Vigileo monitor (Vigileo; Flotrac; Edwards Lifesciences, Irvine, CA, USA) analyzes systemic arterial pressure wave and allows continuous stroke volume monitoring. The aims of this study were (i) to compare changes in stroke volume induced by passive leg raising measured with the Vigileo device and with transthoracic echocardiography and (ii) to compare their ability to predict fluid responsiveness. METHODS: Thirty-four patients with spontaneous breathing activity and considered for volume expansion were included. Measurements of stroke volume were obtained with transthoracic echocardiography (SV-TTE) and with the Vigileo (SV-Flotrac) in a semi-recumbent position, during PLR and after volume expansion (500 ml saline). Patients were responders to volume expansion if SV-TTE increased > or = 15%. RESULTS: Four patients were excluded. No patients received vasoactive drugs. Seven patients presented septic hypovolemia. PLR-induced changes in SV-TTE and in SV-Flotrac were correlated (r2 = 0.56, P < 0.0001). An increase in SV-TTE > or = 13% during PLR was predictive of response to volume expansion with a sensitivity of 100% and a specificity of 80%. An increase in SV-Flotrac > or =16% during PLR was predictive of response to volume expansion with a sensitivity of 85% and a specificity of 90%. There was no difference between the area under the ROC curve for PLR-induced changes in SV-TTE (AUC = 0.96 +/- 0.03) or SV-Flotrac (AUC = 0.92 +/- 0.05). Volume expansion-induced changes in SV-TTE correlated with volume expansion-induced changes in SV-Flotrac (r2 = 0.77, P < 0.0001). In all patients, the highest plateau value of SV-TTE recorded during PLR was obtained within the first 90 s following leg elevation, whereas it was 120 s for SV-Flotrac. CONCLUSIONS: PLR-induced changes in SV-Flotrac are able to predict the response to volume expansion in spontaneously breathing patients without vasoactive support.

Cite

CITATION STYLE

APA

Biais, M., Vidil, L., Sarrabay, P., Cottenceau, V., Revel, P., & Sztark, F. (2009). Changes in stroke volume induced by passive leg raising in spontaneously breathing patients: comparison between echocardiography and Vigileo/FloTrac device. Critical Care (London, England), 13(6). https://doi.org/10.1186/cc8195

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free