Model-based respiratory motion compensation in MRgHIFU

2Citations
Citations of this article
15Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Magnetic Resonance guided High Intensity Focused Ultrasound (MRgHIFU) is an emerging non-invasive technology for the treatment of pathological tissue. The possibility of depositing sharply localised energy deep within the body without affecting the surrounding tissue requires the exact knowledge of the target's position. The cyclic respiratory organ motion renders targeting challenging, as the treatment focus has to be continuously adapted according to the current target's displacement in 3D space. In this paper, a combination of a patient-specific dynamic breath model and a population-based statistical motion model is used to compensate for the respiratory induced organ motion. The application of a population based statistical motion model replaces the acquisition of a patient-specific 3D motion model, nevertheless allowing for precise motion compensation. © 2012 Springer-Verlag.

Cite

CITATION STYLE

APA

Arnold, P., Preiswerk, F., Fasel, B., Salomir, R., Scheffler, K., & Cattin, P. (2012). Model-based respiratory motion compensation in MRgHIFU. In Lecture Notes in Computer Science (including subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics) (Vol. 7330 LNCS, pp. 54–63). https://doi.org/10.1007/978-3-642-30618-1_6