Abstract

Background: As several new tibial osteotomy plates recently appeared on the market, the aim of the present study was to compare mechanical static and fatigue strength of three newly designed plates with gold standard plates for the treatment of medial knee joint osteoarthritis. Methods: Sixteen fourth-generation tibial bone composites underwent a medial open-wedge high tibial osteotomy (HTO) according to standard techniques, using five TomoFix standard plates, five PEEKPower plates and six iBalance implants. Static compression load to failure and load-controlled cyclic fatigue failure tests were performed. Forces, horizontal and vertical displacements were measured; rotational permanent plastic deformations, maximal displacement ranges in the hysteresis loops of the cyclic loading responses and dynamic stiffness were determined. Results: Static compression load to failure tests revealed that all plates showed sufficient stability up to 2400 N without any signs of opposite cortex fracture, which occurred above this load in all constructs at different load levels. During the fatigue failure tests, screw breakage in the iBalance group and opposite cortex fractures in all constructs occurred only under physiological loading conditions (<2400 N). The highest fatigue strength in terms of maximal load and number of cycles performed prior to failure was observed for the ContourLock group followed by the iBalance implants, the TomoFix standard (std) and small stature (sm) plates. The PEEKPower group showed the lowest fatigue strength. Conclusions: All plates showed sufficient stability under static loading. Compared to the TomoFix and the PEEKPower plates, the ContourLock plate and iBalance implant showed a higher mechanical fatigue strength during cyclic fatigue testing. These data suggest that both mechanical static and fatigue strength increase with a wider proximal T-shaped plate design together with diverging proximal screws as used in the ContourLock plate or a closed-wedge construction as in the iBalance design. Mechanical strength of the bone-implant constructs decreases with a narrow T-shaped proximal end design and converging proximal screws (TomoFix) or a short vertical plate design (PEEKPower Plate). Whenever high mechanical strength is required, a ContourLock or iBalance plate should be selected.

References Powered by Scopus

TomoFix: A new LCP-concept for open wedge osteotomy of the medial proximal tibia - Early results in 92 cases

494Citations
N/AReaders
Get full text

Improvements in surgical technique of valgus high tibial osteotomy

494Citations
N/AReaders
Get full text

Tibio-femoral loading during human gait and stair climbing

319Citations
N/AReaders
Get full text

Cited by Powered by Scopus

High Tibial Osteotomy: Review of Techniques and Biomechanics

105Citations
N/AReaders
Get full text

Advances in modern osteotomies around the knee: Report on the Association of Sports Traumatology, Arthroscopy, Orthopaedic surgery, Rehabilitation (ASTAOR) Moscow International Osteotomy Congress 2017

41Citations
N/AReaders
Get full text

The effect of plate design, bridging span, and fracture healing on the performance of high tibial osteotomy plates an experimental and finite element study

34Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Diffo Kaze, A., Maas, S., Waldmann, D., & Zilian, A. (2015). Biomechanical properties of five different currently used implants for open-wedge high tibial osteotomy. Journal of Experimental Orthopaedics, 2(1), 1–17. https://doi.org/10.1186/s40634-015-0030-4

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 23

57%

Researcher 10

25%

Professor / Associate Prof. 6

15%

Lecturer / Post doc 1

3%

Readers' Discipline

Tooltip

Engineering 20

47%

Medicine and Dentistry 19

44%

Nursing and Health Professions 2

5%

Psychology 2

5%

Save time finding and organizing research with Mendeley

Sign up for free