An effective electrochemical biosensing platform for the detection of reduced glutathione

18Citations
Citations of this article
19Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A biosensor was developed using glutathione peroxidase (GSH-Px) as the enzyme. Firstly, platinum (Pt) nanoparticles were deposited onto a glassy carbon paste electrode (GCPE), and then GSH-Px was immobilized by means of gelatin that was then crosslinked with glutaraldehyde. The measurement was based on the electrochemical oxidation of GSH to its disulfide form in the presence of hydrogen peroxide. The linear range was found to be between 10 and 250 μM, with a correlation coefficient of R2 = 0.9968. The R.S.D value for 25 μM GSH (n = 6) was calculated as 2.92%. Finally, the proposed biosensor was used to analyze GSH in a synthetically prepared plasma sample, and a promising recovery value was obtained.

Cite

CITATION STYLE

APA

Anik, Ü., Çubukçu, M., & Ertaş, F. N. (2016). An effective electrochemical biosensing platform for the detection of reduced glutathione. Artificial Cells, Nanomedicine and Biotechnology, 44(3), 971–977. https://doi.org/10.3109/21691401.2015.1008504