A NEW OPEN AND CLOSED RESPIRATION CHAMBER

8Citations
Citations of this article
16Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A new, fully automated, open and closed respiration chamber has been constructed, which is large enough to accommodate man or comparable sized animals. Metabolic rate is calculated from measured changes in respiratory gas volumes while the system is hermetically sealed. When CO2 concentration reaches a prescribed level, large ports are opened and the chamber is ventilated. By repeating this process, a series of metabolic determinations can be conducted over an extended period. Temperature, humidity and air velocity are fixed, but chamber volume and pressure vary with changes in the respiratory exchange of the animal and/or barometric pressure. Small, rapid changes in gas volumes are monitored by a low‐resistance spirometer and large changes accommodated in a second, high‐volume, motorized spirometer. Chamber operations, data aquisition and metabolic calculations are performed by a small computer. System calibration with methanol combustion indicates that estimates of respiratory gas exchange are highly accurate over a temperature range from 10‐40°C. © 1983 The Physiological Society

Cite

CITATION STYLE

APA

Aulick, L. H., Arnhold, H., Hander, E. H., & Mason, A. D. (1983). A NEW OPEN AND CLOSED RESPIRATION CHAMBER. Quarterly Journal of Experimental Physiology, 68(3), 351–357. https://doi.org/10.1113/expphysiol.1983.sp002730