Low specificity of HIV-testing on sputum specimens kept at ambient temperatures for 4 to 7 days: A blinded comparison

3Citations
Citations of this article
20Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background. HIV testing on sputum using the QraQuick HIV1/2® assay has high sensitivity and specificity, and holds promise for application in tuberculosis surveys. Its performance under conditions that may occur during surveys in resource-poor countries is however, unknown. We assessed, in a blinded comparison with HIV serum testing, the sensitivity and specificity of the OraQuick® assay for detecting HIV antibody in sputum specimens kept at ambient temperature for up to 7 days, with and without decontaminant. Methods. Paired sputum and blood specimens from consecutively diagnosed smear-positive tuberculosis patients were tested with OraQuick® and 2 HIV-1/2 ELISA's. Sputum was tested within 24 hours of collection, split into 2 aliquots with and without addition of cetylpyridium chloride, and tested again after 4 and 7 days. Results. Complete data was available for 377/435 (87%) enrolled patients; 132 (35%) tested HIV positive on serum. The sensitivity of the sputum test was 94.7% (95% CI 89.4-97.8) on day 1, 93.2% on day 4 and 92.9% on day 7. The specificity was 92.9% (95% CI 88.9-95.8) on day 1, and declined to 76.7% on day 4 (p < 0.001) and to 62.7% on day 7 (p < 0.001). Adding cetylpyridium chloride further decreased the specificity to 67.8% on day 4 (p = 0.04) and to 49.6% on day 7 (p = 0.004). Conclusion. Transportation of sputum specimens at ambient temperatures for 4 days or more, and addition of decontaminant, strongly affect the specificity of the OraQuick® assay. Unless applied within one day, this assay is not suitable for estimation of HIV-prevalence among tuberculosis patients in survey settings. © 2007 Egwaga et al; licensee BioMed Central Ltd.

Cite

CITATION STYLE

APA

Egwaga, S. M., Chonde, T. M., Matee, M. I., Mfinanga, S. G., Ngowi, P. E., Lwilla, F., & Cobelens, F. G. J. (2007). Low specificity of HIV-testing on sputum specimens kept at ambient temperatures for 4 to 7 days: A blinded comparison. BMC Clinical Pathology, 7. https://doi.org/10.1186/1472-6890-7-8

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free