Absolute quantitative total-body small-animal SPECT with focusing pinholes

36Citations
Citations of this article
44Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Purpose: In pinhole SPECT, attenuation of the photon flux on trajectories between source and pinholes affects quantitative accuracy of reconstructed images. Previously we introduced iterative methods that compensate for image degrading effects of detector and pinhole blurring, pinhole sensitivity and scatter for multi-pinhole SPECT. The aim of this paper is (1) to investigate the accuracy of the Chang algorithm in rodents and (2) to present a practical Chang-based method using body outline contours obtained with optical cameras. Methods: Here we develop and experimentally validate a practical method for attenuation correction based on a Chang first-order method. This approach has the advantage that it is employed after, and therefore independently from, iterative reconstruction. Therefore, no new system matrix has to be calculated for each specific animal. Experiments with phantoms and animals were performed with a high-resolution focusing multi-pinhole SPECT system (U-SPECT-II, MILabs, The Netherlands). This SPECT system provides three additional optical camera images of the animal for each SPECT scan from which the animal contour can be estimated. Results: Phantom experiments demonstrated that an average quantification error of -18.7% was reduced to -1.7% when both window-based scatter correction and Chang correction based on the body outline from optical images were applied. Without scatter and attenuation correction, quantification errors in a sacrificed rat containing sources with known activity ranged from -23.6 to -9.3%. These errors were reduced to values between -6.3 and +4.3% (with an average magnitude of 2.1%) after applying scatter and Chang attenuation correction. Conclusion: We conclude that the modified Chang correction based on body contour combined with window-based scatter correction is a practical method for obtaining small-animal SPECT images with high quantitative accuracy. © 2010 The Author(s).

References Powered by Scopus

A method for attenuation correction in radionuclide computed tomography

1160Citations
N/AReaders
Get full text

A Practical Method for Position-Dependent Compton-Scatter Correction in Single Photon Emission CT

524Citations
N/AReaders
Get full text

Pinhole collimation for ultra-high-resolution, small-field-of-view SPECT

285Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Wu, C., Van Der Have, F., Vastenhouw, B., Dierckx, R. A. J. O., Paans, A. M. J., & Beekman, F. J. (2010). Absolute quantitative total-body small-animal SPECT with focusing pinholes. European Journal of Nuclear Medicine and Molecular Imaging, 37(11), 2127–2135. https://doi.org/10.1007/s00259-010-1519-9

Readers' Seniority

Tooltip

Researcher 16

48%

PhD / Post grad / Masters / Doc 13

39%

Professor / Associate Prof. 4

12%

Readers' Discipline

Tooltip

Physics and Astronomy 7

30%

Medicine and Dentistry 6

26%

Engineering 5

22%

Agricultural and Biological Sciences 5

22%

Save time finding and organizing research with Mendeley

Sign up for free