This is a report on the development of an automated precision load-based measurement setup for thermoelectric power (S) of different types of samples in the temperature range of 5-330 K. The problems in the old spring-based setup have been solved in this load-based setup. This setup takes nearly 4 h for each run, and the typical error is within 5%. High quality calibration has been demonstrated using high purity platinum wires and cylinders.
CITATION STYLE
Tarachand, Saxena, M., Mukherjee, B., & Okram, G. S. (2019). A load-based thermopower measurement setup in the temperature range of 5-330 K. Review of Scientific Instruments, 90(6). https://doi.org/10.1063/1.5090954
Mendeley helps you to discover research relevant for your work.