Introduction Hearing screening allows the identification of individuals with hearing loss. Aim To determine the sensitivity and specificity of a portable hearing screening device in middle-aged and older adults using the manufacturer scoring and a scoring system proposed by the researchers. Methods In this transversal study, participants underwent anamnesis, otoscopy, and hearing screening using portable equipment. After this, a pure tone audiometry was performed, with participants classified into two groups: with and without hearing loss. The sensitivity and specificity of the hearing screening were calculated for the right and left ears using two methods of interpretation: the originalmethod recommended by the manufacturer (criteria 1) and the method proposed by researchers (criteria 2). Results The sample consisted of 55 individuals, 83.6% (n = 46) of whomwere women. Per criteria 1, the sensitivities were 26.3 (right ear) and 21.4% (left ear). The specificity was 100% for both ears. Using criteria 2, the sensitivity was 94.7 (right ear) and 100% (left ear). The specificity was 74.3 (right ear) and 65.9% (left ear). Conclusion This study showed that the criteria proposed by the manufacturer presented low sensitivity in the hearing screening. The criteria proposed by the researchers to achieve a more efficient performance reached high and balanced values for sensitivity and specificity. Copyright © 2014 by Thieme Publicações Ltda, Rio de Janeiro, Brazil.
CITATION STYLE
Cardoso, C. L., Bós, Â. J. G., Gonçalves, A. K., Olchik, M. R., Flores, L. S., Seimetz, B. M., … Teixeira, A. R. (2014). Sensitivity and specificity of portable hearing screening in middle-aged and older adults. International Archives of Otorhinolaryngology, 18(1), 21–26. https://doi.org/10.1055/s-0033-1358660
Mendeley helps you to discover research relevant for your work.