This brief work-in-progress outlines two methods that we have attempted for determining the dose at which the linear relation between optical density of a PRESAGE3 dosimeter and the dose deposited breaks down. Both methods were equally successful in mapping the linear relation up to an optical density of approximately 6.25 cm-1 (absorbance 2..5), but no saturation was found in this region. © 2009 IOP Publishing Ltd.
CITATION STYLE
Al-Nowais, S., Nisbet, A., Adamovics, J., & Doran, S. J. (2009). An attempt to determine the saturation dose for PRESAGEtm. Journal of Physics: Conference Series, 164. https://doi.org/10.1088/1742-6596/164/1/012043