Femoral artery complications associated with the Mynx closure device

30Citations
Citations of this article
17Readers
Mendeley users who have this article in their library.

Abstract

BACKGROUND AND PURPOSE: Devices to close a femoral arteriotomy are frequently used after catheterization for interventional radiology and cardiac procedures to decrease the time to hemostasis and ambulation and, potentially, to decrease local complications. The Mynx vascular closure device uses a sealant designed to occlude the access tract, resulting in hemostasis. MATERIALS AND METHODS: We retrospectively reviewed all cases in which the Mynx device was used and for which follow-up angiography was available. A total of 146 devices were deployed in 135 patients. A follow-up vascular study visualizing the femoral artery was performed in 26 patients (27 studies). RESULTS: There were 5 (5/27, 18%) cases of intravascular Mynx sealant on follow-up vascular imaging. Three pseudoaneurysms (3/27, 11%) were identified. CONCLUSIONS: In this small study, intravascular sealant and pseudoaneurysms were found frequently after femoral arterial closure with the Mynx vascular closure device.

Cite

CITATION STYLE

APA

Fields, J. D., Liu, K. C., Lee, D. S., Gonda, S. J., Dogan, A., Gultekin, S. H., … Barnwell, S. L. (2010). Femoral artery complications associated with the Mynx closure device. American Journal of Neuroradiology, 31(9), 1737–1740. https://doi.org/10.3174/ajnr.A2153

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free