Analysis of polyethylene wear in plain radiographs

  • Stilling M
  • Søballe K
  • Andersen N
 et al. 
  • 12

    Readers

    Mendeley users who have this article in their library.
  • N/A

    Citations

    Citations of this article.

Abstract

Background and purpose Two-dimensional computerized radiographic techniques are frequently used to measure in vivo polyethylene (PE) wear after total hip arthroplasty (THA), and several variables in the clinical set-up may influence the amount of wear that is measured. We compared the repeatability and concurrent validity of linear PE wear on plain radiographs using the same software but a different number of radiographs. Methods We used either 1, 2, or 6 anteroposterior (AP) hip radiographs of 11 patients from a clinical THA series with 12 years of follow-up, and measured the PE wear with the software PolyWare 3D Pro. Repeatability within and concurrent validity between the different numbers of radiograph strategies were assessed using limits of agreement (LOAs) and bias. Results Observed median wear (range) in mm was 3.4 (1.64.6), 2.3 (0.74.9), and 4.0 (2.66.2) for the 1-, 2-, and 6-radiograph strategies. For repeatability, no bias (p > 0.41) was observed. LOAs around the bias were 0.6, 0.4, and 1.2 mm for the 1-, 2-, and 6-radiograph strategies. For concurrent validity, a bias LOA) between all pairwise comparisons was observed (p < 0.02) with 0.8 mm 2.5) between the 1- and 2-radiograph strategies, 1.0 mm 2.2) between the 1- and 6-radiograph strategies, and 1.8 mm 1.2) between the 2- and 6-radiograph strategies. Interpretation The number of radiographs used for wear measurement with a shadow-casting analysis method on plain AP radiographs influences the amount of linear wear measured. Results of PE wear obtained with PolyWare in studies using a different number of radiographs are not comparable.

Author-supplied keywords

  • arthroplasty
  • compressive strength
  • equipment failure analysis
  • femur head
  • femur head radiography
  • follow up studies
  • hip
  • hip adverse effects
  • hip joint
  • hip joint radiography
  • hip prosthesis
  • hip prosthesis adverse effects
  • humans
  • imaging
  • mechanical
  • observer variation
  • polyethylene
  • prosthesis design
  • prosthesis failure
  • reoperation
  • replacement
  • reproducibility results
  • software
  • stress
  • three dimensional

Get free article suggestions today

Mendeley saves you time finding and organizing research

Sign up here
Already have an account ?Sign in

Authors

  • Maiken Stilling

  • Kjeld Søballe

  • Niels Trolle Andersen

  • Kristian Larsen

  • Ole Rahbek

Cite this document

Choose a citation style from the tabs below

Save time finding and organizing research with Mendeley

Sign up for free