Coagulation-induced resistance to fluid flow in small-diameter vascular grafts and graft mimics measured by purging pressure

  • Nichols M
  • Choudhary R
  • Kodali S
 et al. 
  • 8

    Readers

    Mendeley users who have this article in their library.
  • 2

    Citations

    Citations of this article.

Abstract

In this study, the coagulation-induced resistance to flow in small-diameter nonpermeable Tygon tubes and permeable expanded polytetrafluoroethylene (ePTFE) vascular grafts was characterized by measuring the upstream pressure needed to purge the coagulum from the tube lumen. This purging pressure was monitored using a closed system that compressed the contents of the tubes at a constant rate. The pressure system was validated using a glycerin series with well-defined viscosities and precisely controlled reductions in cross-sectional area available for flow. This system was then used to systematically probe the upstream pressure buildup as fibrin glue, platelet-rich plasma (PRP) or whole blood coagulated in small-diameter Tygon tubing and or ePTFE grafts. The maximum purging pressures rose with increased clot maturity for fibrin glue, PRP, and whole blood in both Tygon and ePTFE tubes. Although the rapidly coagulating fibrin glue in nonpermeable Tygon tubing yielded highly consistent purging curves, the significantly longer and more variable clotting times of PRP and whole blood, and the porosity of ePTFE grafts, significantly diminished the consistency of the purging curves.

Author-supplied keywords

  • fibrin
  • hemocompatibility
  • test method
  • thrombogenicity
  • vascular graft

Get free article suggestions today

Mendeley saves you time finding and organizing research

Sign up here
Already have an account ?Sign in

Find this document

Authors

  • Michael D. Nichols

  • Rewa Choudhary

  • Santhisri Kodali

  • William M. Reichert

Cite this document

Choose a citation style from the tabs below

Save time finding and organizing research with Mendeley

Sign up for free