Conventional 2D-EPI or Segmented 3D-EPI? A Temporal SNR Study at 3 and 7 Tesla

  • Stirnberg R
  • Stöcker T
N/ACitations
Citations of this article
17Readers
Mendeley users who have this article in their library.

Abstract

This work studies the temporal SNR characteristics of conventional 2D-EPI and segmented 3D-EPI acquisitions under realistic conditions for standard and advanced applications. Therefore, different whole-brain protocols optimized for BOLD fMRI at normal and high spatial resolutions are prepared. Time series data are acquired at 3 and 7 Tesla. As expected, segmented 3D-EPI achieves higher tSNR than 2D-EPI at high spatial resolutions (thermal noise dominated regime). Beyond that, however, a general signal sensitivity advantage due to higher imaging acceleration capabilities (larger data rate, hence increased robustness of fMRI data) renders 3D-EPI useful even at typical, rather coarse resolutions.

Cite

CITATION STYLE

APA

Stirnberg, R., & Stöcker, T. (2014). Conventional 2D-EPI or Segmented 3D-EPI? A Temporal SNR Study at 3 and 7 Tesla. In Proc. Intl. Soc. Mag. Reson. Med. 22 (Vol. 22, p. 0868).