Derivation of sonograph quality parameters by the use of point spread function analysis

3Citations
Citations of this article
7Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A measuring system evaluating a Point Spread Function generated in an ultrasonographic image by scanning a spherical target was developed. The target is moved in measuring bath filled by water over scanned volume via 3D computer controlled positioning system. A video signal obtained is converted to digital form and analyzed by original software to derive various objective parameters of the imager as follows: Focal areas in both the azimuth and the elevation directions, Ultrasound scanning lines visualisation, Manufacturer preloaded TGC, Width of the scanning plane, Side lobe levels and Amplification uniformity in the azimuth direction. The method was verified by testing 18 different equipments in 282 measurements. Samples of particular measurement results in form of graphical outputs are included. Medical and physiological impacts of this approach are discussed. © 2007 Institute of Physiology v.v.i., Academy of Sciences of the Czech Republic.

Cite

CITATION STYLE

APA

Doležal, L., Mazura, J., Tesařík, J., Kolářová, H., Korpas, D., Binder, S., & Hálek, J. (2007). Derivation of sonograph quality parameters by the use of point spread function analysis. Physiological Research, 56(SUPPL. 1). https://doi.org/10.33549/physiolres.931304