Independence of myoelectric control signals examined using a surface EMG model

46Citations
Citations of this article
68Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The detection volume of the surface electromyographic (EMG) signal was explored using a finite-element model, to examine the feasibility of obtaining independent myoelectric control signals from regions of reinnervated muscle. The selectivity of the surface EMG signal was observed to decrease with increasing subcutaneous fat thickness. The results confirm that reducing the interelectrode distance or using double-differential electrodes can increase surface EMG selectivity in an inhomogeneous volume conductor. More focal control signals can be obtained, at the expense of increased variability, by using the mean square value, rather than the root mean square or average rectified value.

Cite

CITATION STYLE

APA

Lowery, M. M., Stoykov, N. S., & Kuiken, T. A. (2003). Independence of myoelectric control signals examined using a surface EMG model. IEEE Transactions on Biomedical Engineering, 50(6), 789–793. https://doi.org/10.1109/TBME.2003.812152