Investigating the range of surgical effects on soft tissue produced by a carbon dioxide laser

25Citations
Citations of this article
12Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The authors investigated the surgical and collateral effects on soft tissue of a carbon dioxide laser emitting at 9.3 micrometers. Specifically, incision widths and depths as well as effectiveness were studied. Three different laser modes were investigated: gated continuous wave, or Cw, Superpulse and OptiPulse (Medical Optics). Incision depths correlated positively with average power; higher powers produced deeper incisions. The gated Cw mode quickly produced wide, deep incisions; Superpulse achieved narrower, deep incisions; OptiPulse caused very narrow, shallow incisions. Collateral damage to adjacent tissues was reduced by a factor of about 2 using Superpulse, and by a factor of 10 using OptiPulse. A wide range of effects is achieved in soft tissue, depending on the laser parameter combination used.

Cite

CITATION STYLE

APA

Wilder-Smith, P., Dang, J., & Kurosaki, T. (1997). Investigating the range of surgical effects on soft tissue produced by a carbon dioxide laser. Journal of the American Dental Association, 128(5), 583–588. https://doi.org/10.14219/jada.archive.1997.0257