Rapid method for the measurement of major and minor elements in vanadium titanium magnetite sample

ISSN: 10000593
1Citations
Citations of this article
N/AReaders
Mendeley users who have this article in their library.

Abstract

A high resolution (203 eV/Mn Kα) portable X-ray fluorescence spectrometer has been developed and the measurement of major and minor elements in vanadium titanium magnetite sample was carried out. An 241Am radioactive source for X-ray excitation, a Si-PIN detector and a 2048 channel analyzer for the measurement of vanadium titanium magnetite direct powder sample were used. The correction methods for interferences, calibration and drift are presented. The method is simply, accurate and rapid. The detection limit is an order of magnitude lower compared with a sealed proportional counter system.