Recommendations on the practice of calibration, dosimetry, and quality assurance for gamma stereotactic radiosurgery: Report of AAPM Task Group 178

25Citations
Citations of this article
47Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The American Association of Physicists in Medicine (AAPM) formed Task Group 178 (TG-178) to perform the following tasks: review in-phantom and in-air calibration protocols for gamma stereotactic radiosurgery (GSR), suggest a dose rate calibration protocol that can be successfully utilized with all gamma stereotactic radiosurgery (GSR) devices, and update quality assurance (QA) protocols in TG-42 (AAPM Report 54, 1995) for static GSR devices. The TG-178 report recommends a GSR dose rate calibration formalism and provides tabulated data to implement it for ionization chambers commonly used in GSR dosimetry. The report also describes routine mechanical, dosimetric, and safety checks for GSR devices, and provides treatment process quality assurance recommendations. Sample worksheets, checklists, and practical suggestions regarding some QA procedures are given in appendices. The overall goal of the report is to make recommendations that help standardize GSR physics practices and promote the safe implementation of GSR technologies.

Cite

CITATION STYLE

APA

Petti, P. L., Rivard, M. J., Alvarez, P. E., Bednarz, G., Daniel Bourland, J., DeWerd, L. A., … Goetsch, S. J. (2021, July 1). Recommendations on the practice of calibration, dosimetry, and quality assurance for gamma stereotactic radiosurgery: Report of AAPM Task Group 178. Medical Physics. John Wiley and Sons Ltd. https://doi.org/10.1002/mp.14831

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free