Scatter measurements for optical cone-beam computed tomography

5Citations
Citations of this article
11Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Both x-ray and optical cone-beam CT (CBCT) scanners are limited by scattered photons that contaminate the primary images. Transparent radiochromic gels and plastics are anticipated to produce less scattering than polymerization gels and hence result in reconstructed dose distributions with greater contrast and dynamic range. Four methods of scatter measurement were investigated to quantify scatter fractions using an in-house optical CBCT scanner. The methods consisted of generating either "coin" shadows or primary beamlets in transmission images. The four methods generated similar results of 10% scatter fractions in gel transmission images for this particular scanner configuration with both leuco crystal violet and carbon black micelle gels. The scatter contributions were distributed as 5% from micelle gel, 2% from Teflon wall, 2% from refractive index matching liquid and 1% other sources. Applying a uniform background subtraction of 2% of open field intensity resulted in 0.2% difference between laser CT and cone-beam CT reconstructed attenuation coefficients for a black micelle gel finger phantom. © 2009 IOP Publishing Ltd.

Cite

CITATION STYLE

APA

Jordan, K., & Battista, J. (2009). Scatter measurements for optical cone-beam computed tomography. Journal of Physics: Conference Series, 164. https://doi.org/10.1088/1742-6596/164/1/012028

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free