A cement spacer for two-stage revision of infected implants of the hip joint

125Citations
Citations of this article
39Readers
Mendeley users who have this article in their library.

Abstract

We report the technical details and clinical results of twelve patients who had deep infections of implants in the hip joint and were treated by two-stage revision, using a gentamicin-loaded, hand-moulded cement spacer inserted for the period between resection and reimplantation arthroplasty. During management with the spacer, usually for 4 months, patients were almost free of pain and mobile with good leg control, spending 2/3 of the treatment period at home. Six of twelve spacers failed locally due to dislocation [5] or cement fracture [1], and more than two further episodes of surgery were required in 3 patients. Problems with dislocation of the spacer were significantly higher when the head to neck offset was lacking (P < 0.05) or when anchorage in the femoral shaft was poor. Nevertheless, infection after reimplantation arthroplasty did not occur by the time of follow-up (2.2 years). Based on these data, we consider that the use of the cement spacer is a promising approach to the treatment of complicated infections of the hip joint.

Cite

CITATION STYLE

APA

Leunig, M., Chosa, E., Speck, M., & Ganz, R. (1998). A cement spacer for two-stage revision of infected implants of the hip joint. International Orthopaedics, 22(4), 209–214. https://doi.org/10.1007/s002640050244