Abstract
A novel method of determining trace manganese by square-wave stripping voltammetry with simultaneous plating mercury electrode is described in this paper. Well-defined stripping peaks of manganese are obtained in H 3BO4-NaOH buffer (pH 8.0) and peak currents are in good linear relationship with manganese concentrations in two ranges: from 0.36 nM to 36 nM and from 73 nM to 909 nM. The obtained detection limit is 0.18 nM and relative standard deviation is 2.3% (n=7) under a predeposition time of 90 s. This proposed method has been used for the determination of trace manganese in real plasma samples with satisfiable results and good matching with the results of graphite furnace atomic absorption spectrometry. © 2008 Wiley-VCH Verlag GmbH & Co. KGaA.
Author supplied keywords
Cite
CITATION STYLE
Wang, S., & Ye, B. (2008). Determination of trace manganese by square-wave stripping voltammetry. Electroanalysis, 20(9), 984–988. https://doi.org/10.1002/elan.200704136