Abstract
The Laser Fusion Experiments Groups from the Laboratory for Laser Energetics (LLE) and the Los Alamos National Laboratory (LANL) have jointly developed an instrument capable of simultaneously space-, time-, and spectrally resolving x-ray emission from inertial confinement fusion (ICF) targets. Uses of the instrument include framed imaging of line emission from fuel or shell dopants and monochromatic backlighting. The x-ray imaging is accomplished with a Kirkpatrick-Baez (KB)-type four-image microscope, which has a best spatial resolution of ∼5 μm and a sensitive energy range of ∼2-8 keV. Time-resolved x-ray images are obtained with a pair of custom framing cameras, each of which records two of the four images in two independent 80-ps time intervals. In addition, the energy range of the images can be restricted to a narrow (monochromatic) spectral range (∼10-100 eV) by the introduction of diffracting crystals. This technique has been demonstrated with an e-beam-generated dc x-ray source, and at the LANL Trident laser facility and the LLE OMEGA laser facility with x-rays from laser-produced plasmas. © 1997 American Institute of Physics.
Cite
CITATION STYLE
Marshall, F. J., & Oertel, J. A. (1997). A framed monochromatic x-ray microscope for ICF (invited). Review of Scientific Instruments, 68(1), 735–739. https://doi.org/10.1063/1.1147688