Reducing the impact of wind noise on cochlear implant processors with two microphones

  • Kokkinakis K
  • Cox C
4Citations
Citations of this article
31Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Behind-the-ear (BTE) processors of cochlear implant (CI) devices offer little to almost no protection from wind noise in most incidence angles. To assess speech intelligibility, eight CI recipients were tested in 3 and 9 m/s wind. Results indicated that speech intelligibility decreased substantially when the wind velocity, and in turn the wind sound pressure level, increased. A two-microphone wind noise suppression strategy was developed. Scores obtained with this strategy indicated substantial gains in speech intelligibility over other conventional noise reduction strategies tested.

Cite

CITATION STYLE

APA

Kokkinakis, K., & Cox, C. (2014). Reducing the impact of wind noise on cochlear implant processors with two microphones. The Journal of the Acoustical Society of America, 135(5), EL219–EL225. https://doi.org/10.1121/1.4871583