Novel fully-integrated biosensor for endotoxin detection via polymyxin B immobilization onto gold electrodes

9Citations
Citations of this article
23Readers
Mendeley users who have this article in their library.

Abstract

In this paper an electrochemical endotoxin biosensor consisting of an immobilized lipopolysaccharide (LPS) ligand, polymyxin B (PmB), is presented. Several parameters involved both in the device fabrication and in the detection process were analyzed to optimize the ligand immobilization and the interaction between PmB and LPS, aiming at increasing the sensitivity of the sensor. Different electrochemical pre-treatment procedures as well as the functionalization methods were studied and evaluated. The use of a SAM (self-assembled monolayer) to immobilize PmB and the quantification of the interactions via cyclic voltammetry allowed the development of a robust and simple device for in situ detection of LPS. Thus, the biosensor proposed in this work intends an approach to the demanding needs of the market for an integrated, portable and simple instrument for endotoxin detection.

Cite

CITATION STYLE

APA

Zuzuarregui, A., Arana, S., Pérez-Lorenzo, E., Sánchez-Gómez, S., Martínez De Tejada, G., & Mujika, M. (2013). Novel fully-integrated biosensor for endotoxin detection via polymyxin B immobilization onto gold electrodes. Journal of Sensors and Sensor Systems, 2(2), 157–164. https://doi.org/10.5194/jsss-2-157-2013

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley