Prospective multiaxial motion correction for fMRI

114Citations
Citations of this article
119Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Corruption of the image time series due to interimage head motion limits the clinical utility of functional MRI. This paper presents a method for real-time prospective correction of rotation and translation in all six degrees of rigid body motion. By incorporating an orbital navigator (ONAV) echo for each of the sagittal, axial, and coronal planes into the fMRI pulse sequence, rotation and translation can be measured and the spatial orientation of the image acquisition sequence that follows can be corrected prospectively in as little as 160 msec. Testing of the method using a computerized motion phantom capable of performing complex multiaxial motion showed subdegree rotational and submillimeter translational accuracy over a range of ±8°and ±8 mm of motion. In vivo images demonstrate correction of simultaneous through-plane and in-plane motion and improved detection of fMRI activation in the presence of head motion. (C) 2000 Wiley-Liss, Inc.

Cite

CITATION STYLE

APA

Ward, H. A., Riederer, S. J., Grimm, R. C., Ehman, R. L., Felmlee, J. P., & Jack, C. R. (2000). Prospective multiaxial motion correction for fMRI. Magnetic Resonance in Medicine, 43(3), 459–469. https://doi.org/10.1002/(SICI)1522-2594(200003)43:3<459::AID-MRM19>3.0.CO;2-1

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free