Abstract

FRONTier RADiography is an R&D project to assess the feasibility of digital mammography with Synchrotron Radiation at the ELETTRA Light Source in Trieste. In order to reach an acceptable time duration of the exam, a fast- and low-noise photon counting ASIC has been developed in collaboration with Aurelia Microelettronica, called Frontrad ReadOut SysTem. It is a multichannel counting system, each channel being made of a low-noise charge-sensitive preamplifier optimized for X-ray energy range (10-100 keV), a CR-RC2 shaper, a discriminator and a 16-bit counter. In order to set the discriminator threshold, a set of a global 6-bit DAC and a local (per channel) 3-bit DAC has been implemented within the ASIC. We report on the measurements done with the 8-channel prototype chip and the comparison with the simulation results. © 2001 Elsevier Science B.V.

Cite

CITATION STYLE

APA

Prest, M., Vallazza, E., Chiavacci, M., Mariani, R., Motto, S., Neri, M., … Castelli, E. (2001). FROST: A low-noise high-rate photon counting ASIC for X-ray applications. Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 461(1–3), 435–439. https://doi.org/10.1016/S0168-9002(00)01268-7