Intrinsic optical properties of Boston Keratoprosthesis

4Citations
Citations of this article
18Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Purpose: To benchmark the optical performance of Boston Keratoprosthesis (B-KPro). Methods: Back focal lengths (BFL) of B-KPros for various eye axial lengths were measured using an optical bench, International Organization for Standardization– certified for intraocular lens characterization, and compared against manufacturer’s specification. The modulation transfer function (MTF) and the resolution efficiencies were measured. The theoretical geometry-dependent higher-order aberrations (HOA) were calculated. The devices were characterized with optical profilometry for estimating the surface scattering. Aberration correction and subsequent image quality improvement were simulated in CODE-V. Natural scene-imaging was performed in a mock ocular environment. Retrospective analysis of 15 B-KPro recipient eyes were presented to evaluate the possibility of achieving 20/20 best-corrected visual acuity (BCVA). Results: BFL measurements were in excellent agreement with the manufacturer-reported values (r = 0.999). The MTF specification exceeded what is required for achieving 20/20 visual acuity. Astigmatism and field curvature, correctable in simulations, were the primary aberrations limiting imaging performance. Profilometry of the anterior surface revealed nanoscale roughness (root-mean-square amplitude, 30–50 nm), contributing negligibly to optical scattering. Images of natural scenes obtained with a simulated B-KPro eye demonstrated good central vision, with 10/10 visual acuity (equiv-alent to 20/20). Full restoration of 20/20 BCVA was obtainable for over 9 years in some patients. Conclusions: Theoretical and experimental considerations demonstrate that B-KPro has the optical capacity to restore 20/20 BCVA in patients. Further image quality improvement can be anticipated through correction of HOAs. Translational Relevance: We establish an objective benchmark to characterize the optics of the B-KPro and other keratoprosthesis and propose design changes to allow improved vision in B-KPro patients.

Cite

CITATION STYLE

APA

Hui, P. C., Pereira, L. A., Dore, R., Chen, S., Taniguchi, E., Chodosh, J., … Paschalis, E. I. (2020). Intrinsic optical properties of Boston Keratoprosthesis. Translational Vision Science and Technology, 9(12), 1–13. https://doi.org/10.1167/tvst.9.12.10

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free