Monte carlo modeling of a novalis TX Varian 6 MV with HD-120 multileaf collimator

13Citations
Citations of this article
51Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A Monte Carlo model of the Novalis Tx linear accelerator equipped with high-definition multileaf collimator (HD-120 HD-MLC) was commissioned using ionization chamber measurements in water. All measurements in water were performed using a liquid filled ionization chamber. Film measurements were made using EDR2 film in solid water. Open rectangular fields defined by the jaws or the HD-MLC were used for comparison against measurements. Furthermore, inter- and intraleaf leakage calculated by the Monte Carlo model was compared against film measurements. The statistical uncertainty of the Monte Carlo calculations was less than 1% for all simulations. Results for all regular field sizes show an excellent agreement with commissioning data (percent depth-dose curves and profiles), well within 1% of difference in the relative dose and 1 mm distance to agreement. The computed leakage through HD-MLCs shows good agreement with film measurements. The Monte Carlo model developed in this study accurately represents the new Novalis Tx Varian linac with HD-MLC and can be used for reliable patient dose calculations.

Cite

CITATION STYLE

APA

Vazquez-Quino, L. A., Massingill, B., Shi, C., Gutierrez, A., Esquivel, C., Eng, T., … Stathakis, S. (2012). Monte carlo modeling of a novalis TX Varian 6 MV with HD-120 multileaf collimator. Journal of Applied Clinical Medical Physics, 13(5), 300–313. https://doi.org/10.1120/jacmp.v13i5.3960

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free