Analysis of "accuracy and User Performance Evaluation of a New, Wireless-Enabled Blood Glucose Monitoring System That Links to a Smart Mobile Device"

1Citations
Citations of this article
10Readers
Mendeley users who have this article in their library.

Abstract

In the study published in Journal of Diabetes, Science and Technology, Timothy S. Bailey and coauthors evaluated system accuracy and user performance of a new, wireless-enabled blood glucose monitoring system, the Contour Plus ONE blood glucose monitoring system. The authors declare that the study protocol was performed according to ISO 15197:2013 guidelines, in particular to clauses 6.3 and 8. Results show a high level of accuracy of the product in a laboratory setting as well as in the hands of lay users. However, differences between both study parts emerge that stimulate discussion about the used comparison method.

Cite

CITATION STYLE

APA

Rutschmann, M. (2017). Analysis of “accuracy and User Performance Evaluation of a New, Wireless-Enabled Blood Glucose Monitoring System That Links to a Smart Mobile Device.” Journal of Diabetes Science and Technology, 11(4), 744–745. https://doi.org/10.1177/1932296817691306