Abstract

In a laboratory model, humidity output was investigated in relation to the design of a circle absorber system. A 70-kg subject was simulated with fresh gas flows of 0.5, 2 or 5 litre min-1. Different circle systems, absorption canisters and tubings were studied. It was found that both the type of circle system (according to Eger's classification) and the size of the absorption canister, influenced the humidity of inspired gases. Coaxial tubing only moderately increased the humidity. If a fresh gas flow of 0.5 litre min-1 was used, optimum moisture contents were attained, irrespective of the circle system tested. Low fresh gas flows, a small canister and an Eger A type circle system, were factors which increased humidification. © 1989 British Journal of Anaesthesia.

Cite

CITATION STYLE

APA

Bengtson, J. P., Bengtson, A., & Stenqvist, O. (1989). The circle system as a humidifier. British Journal of Anaesthesia, 63(4), 453–457. https://doi.org/10.1093/bja/63.4.453