Optic disc measurement: A comparison of indirect ophthalmoscopic methods

25Citations
Citations of this article
22Readers
Mendeley users who have this article in their library.

Abstract

Aims-Two methods of indirect ophthalmoscopic estimation of optic disc size, the 78 dioptre lens and optic disc biometer were evaluated. Methods-Twenty nine eyes of 29 patients were measured by both methods and compared with optic disc size calculated using the three planimetric corrections described by Bengtsson and Krakau. Results-The closest agreement with the clinical measurements was found using correction 3. There was a significant difference between both the 78 D lens (p<0.0001) and the biometer (p=0.0027) and the planimetric results. There was also a significant difference between the two clinical methods (p<0.0001). Both methods showed acceptable intraobserver variation (CoV 2.45% and 3.13% respectively). Conclusion-Overall, both methods give larger measurements than planimetry; the 78 D lens by 0.41 mm and the biometer by 0.15 mm. Neither method gives a satisfactory estimation of optic disc size when compared with planimetry.

Cite

CITATION STYLE

APA

Spencer, A. F., & Vernon, S. A. (1995). Optic disc measurement: A comparison of indirect ophthalmoscopic methods. British Journal of Ophthalmology, 79(10), 910–915. https://doi.org/10.1136/bjo.79.10.910