NMR SHIM coil design utilising a rapid spherical harmonic calculation method

11Citations
Citations of this article
32Readers
Mendeley users who have this article in their library.

Abstract

A rapid spherical harmonic calculation method is used for the design of Nuclear Magnetic Resonance shim coils. The aim is to design each shim such that it generates a field described purely by a single spherical harmonic. By applying simulated annealing techniques, coil arrangements are produced through the optimal positioning of current-carrying circular arc conductors of rectangular cross-section. This involves minimizing the undesirable harmonics in relation to a target harmonic. The design method is flexible enough to be applied for the production of coil arrangements that generate fields consisting significantly of either zonal or tesseral harmonics. Results are presented for several coil designs which generate tesseral harmonics of degree one. © Australian Mathematical Society 2002.

Cite

CITATION STYLE

APA

Snape-Jenkinson, C. J., Crozier, S., & Forbes, L. K. (2002). NMR SHIM coil design utilising a rapid spherical harmonic calculation method. ANZIAM Journal, 43(3), 375–386. https://doi.org/10.1017/s1446181100012578