This case report describes a rare case of a broken periodontal probe tip and its removal. A male patient presented himself in June 2019 due to a painful tooth in the upper left quadrant. The patient elected treatment in the dental school's student course. In October 2019, in preparation for full-mouth rehabilitation, a complete diagnostic status was performed, including radiographs. In this context, a metal-dense fragment was identified in the apical region of the (missing) tooth 45. It was diagnosed as the broken tip of a periodontal probe (type AE P OWB). Since a PCP-12 probe is generally used in-house, iatrogenic damage during the initial examination or student course could be excluded a priori. The patient was not able to remember any treatment that could be associated with the instrument's breaking. Since the probe fragment was palpable and a translocation could not be precluded, the patient agreed to its removal under local anesthesia, after a cone-beam CT. This article describes and discusses this particular case, with special emphasis on iatrogenic instrument fractures and their removal.
Author supplied keywords
Cite
CITATION STYLE
Kaufmann, M. E., Solderer, A., Hofer, D., & Schmidlin, P. R. (2020). The Strange Case of a Broken Periodontal Instrument Tip. Dentistry Journal, 8(2). https://doi.org/10.3390/DJ8020055