Abstract
Objectives: The aim of this study was to compare the value of the AI-SONIC ultrasound-assisted diagnosis system versus contrast-enhanced ultrasound (CEUS) for differential diagnosis of thyroid nodules in diffuse and non-diffuse backgrounds. Methods: A total of 555 thyroid nodules with pathologically confirmed diagnosis were included in this retrospective study. The diagnostic efficacies of AI-SONIC and CEUS for differentiating benign from malignant nodules in diffuse and non-diffuse backgrounds were evaluated, with pathological diagnosis as the gold standard. Results: The agreement between AI-SONIC diagnosis and pathological diagnosis was moderate in diffuse backgrounds (κ = 0.417) and almost perfect in non-diffuse backgrounds (κ = 0.81). The agreement between CEUS diagnosis and pathological diagnosis was substantial in diffuse backgrounds (κ = 0.684) and moderate in non-diffuse backgrounds (κ = 0.407). In diffuse backgrounds, AI-SONIC had slightly higher sensitivity (95.7 vs 89.4%, P =.375), but CEUS had significantly higher specificity (80.0 vs 40.0%, P =.008). In non-diffuse background, AI-SONIC had significantly higher sensitivity (96.2 vs 73.4%, P
Author supplied keywords
Cite
CITATION STYLE
Liu, T., Wu, C., Wang, G., Jia, Y., Zhu, Y., & Nie, F. (2023). Clinical Value of Artificial Intelligence-Based Computer-Aided Diagnosis System Versus Contrast-Enhanced Ultrasound for Differentiation of Benign From Malignant Thyroid Nodules in Different Backgrounds. Journal of Ultrasound in Medicine, 42(8), 1757–1766. https://doi.org/10.1002/jum.16195