We have evaluated three analytical methods for determining inorganic phosphorus in serum applied to the Technicon RA-1000 analyzer: a fully enzymatic colorimetric method based on the specific system purine nucleoside Phosphorylase/xanthine oxidase coupled to an indicator colorimetric reaction similar to the Trinder reaction; a chemical method involving the direct UV measurement of the phosphomolybdate complex; and a chemical method with reduction of the phosphomolybdate complex to molybdenum blue. Experiments were performed to assess within-run and between-day precision, linearity, interference and correlation. The best performance characteristics were shown by the enzymatic colorimetric method and the phosphomolybdate UV method. © 1989, Hindawi Publishing Corporation. All rights reserved.