Feasibility of PET/CT 3-D dosimetry for proton-activated PRESAGE® dosimeters

1Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A feasibility study is considered using PRESAGE® dosimeters as a direct dosimetry comparison tool to correlate proton activation to dose in a clinical beam. Proton activation measured by PET/CT has been studied as a future tool method for in vivo dosimetry. PRESAGE® is a 3D offline dosimeter that has a material composition giving it a longer activation decay time than similar dosimeters allowing easier measurements. In this study, the PRESAGE® positron emissions were measured by PET and directly compared to dose measurements made by optical-CT to make spatial comparisons of dose and proton activation. Profiles along the central axis of the beam found a shift in the distal-80% resulting in the proton activation profile being 1 cm shallower than the dose profile. Cross beam profiles demonstrated little discrepancy between long (3 hour) and short (30 minute) PET scan times.

Cite

CITATION STYLE

APA

Carroll, M., Ibbott, G., & Adamovics, J. (2015). Feasibility of PET/CT 3-D dosimetry for proton-activated PRESAGE® dosimeters. In Journal of Physics: Conference Series (Vol. 573). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/573/1/012071