Abstract
The aim of this study was to assess whether slanted gel pads can be used to optimize beam-flow angles and flow velocity measurements for Doppler ultrasound. The right carotid artery of a single healthy female volunteer was measured alternatively five times without and with an 18° angled slanted gel pad between the ultrasound transducer and skin by 13 radiologists. Beam-flow angles and peak systolic flow velocities (PSV) were measured along with assessment of spectral broadening. Beam-flow angles (P = 0.001) and PSV (P = 0.001) measurements showed a significant decrease when using slanted gel pads. The mean (±SD) beam-flow angles without and with the use of slanted gel pads were 66.7 (±4.2) and 56.1 (±5.8) degrees, respectively. The mean (±SD) PSVs without and with the use of slanted gel pads were 92.0 (±17.4) and 76.9 (±10.9) cm/s, respectively. There was a noticeable decrease in spectral broadening when using slanted gel pads. There was a significant linear correlation between beam-flow angle and peak systolic velocity. Coefficients of variation for peak systolic velocity without and with the use of gel pads were 18.9 and 14.2 %, respectively. These results demonstrate that slanted gel pads decrease beam-flow angles and overestimation of Doppler flow velocity measurements while potentially increasing the reliability of measurements.