Abstract
This paper presents the development of a cheap and rapid electrochemical sensor for ascorbic acid detection. In particular, the graphite ink working electrode of screen-printed cells was covered by a film of elecrosynthesized molecularly imprinted polypyrrole (e-MIP); differential pulse voltammetry (DPV) was the selected method for the analyte detection. The ascorbic acid molecules were successfully entrapped in the polypyrrole film, creating the recognition sites. The best results were obtained after polypyrrole overoxidation and performing the measurements in phosphate buffer solution 0.05 M/KCl 0.1 M at pH 7.5. A comparison with the bare and the not-imprinted polypyrrole-modified electrodes showed that the e-MIP-based sensor had the highest selectivity and reproducibility. The developed method was applied to assess ascorbic acid in farmaceutical products, obtaining values not significantly different from the declared content.
Author supplied keywords
Cite
CITATION STYLE
Alberti, G., Zanoni, C., Magnaghi, L. R., & Biesuz, R. (2023). Ascorbic Acid Sensing by Molecularly Imprinted Electrosynthesized Polymer (e-MIP) on Screen-Printed Electrodes. Chemosensors, 11(6). https://doi.org/10.3390/chemosensors11060348
Register to see more suggestions