Auditory brainstem responses to chirps delivered by an insert earphone with equalized frequency response

  • Elberling C
  • Don M
  • Kristensen S
10Citations
Citations of this article
41Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Recently it has been demonstrated that auditory brainstem responses, ABRs, to chirps are larger with the ER-2 than with the ER-3A insert earphone due to differences between the corresponding amplitude-frequency responses. Therefore a modified chirp, which equalizes the amplitude-frequency response of the ER-3A, is constructed and subsequently compared to the unmodified chirp. ABRs are recorded from 20 normal-hearing subjects in response to the two chirps delivered by the ER-3A earphone at a wide range of levels. The results confirm that the modified chirp generates significantly larger ABRs than the unmodified chirp at levels below 60 dB nHL.

Cite

CITATION STYLE

APA

Elberling, C., Don, M., & Kristensen, S. G. B. (2012). Auditory brainstem responses to chirps delivered by an insert earphone with equalized frequency response. The Journal of the Acoustical Society of America, 132(2), EL149–EL154. https://doi.org/10.1121/1.4737915