Quality control of cathode-ray tube monitors for medical imaging using a simple photometer

18Citations
Citations of this article
11Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

As computer monitors are used more in medical imaging and the use of picture archiving and communication system workstations with multiple monitors is increasing, quality control protocols become necessary to track subtle variations in performance characteristics. Several tests based on previously published work were applied to 10 monitors of three different types over a period of 5 months. Each test is explained, and the results are shown. For example, without corrective adjustments, 2 monitors from the same workstation showed a small but steady decline in maximum luminance of 8.1% and 7.6% over the course of 11 weeks that was not perceptible. From this experience, we took the first step toward developing a practical and useful quality control protocol. The proposed protocol requires only a photometer and the ability to generate the Society of Motion Picture and Television Engineers (SMPTE) test pattern, and takes approximately 5 minutes per monitor per week to gather data. © 1995 Society for Imaging Informatics in Medicine.

Cite

CITATION STYLE

APA

Parsons, D. M., Kim, Y., & Haynor, D. R. (1995). Quality control of cathode-ray tube monitors for medical imaging using a simple photometer. Journal of Digital Imaging, 8(1), 10–20. https://doi.org/10.1007/BF03168051