Novel device accurately measures graft resistance and compliance to ensure quality of coronary artery bypass

0Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

Abstract

Introduction: The purpose of this study is to know the influence of coronary artery bypass grafting (CABG) on coronary circulation. In the present study, we evaluated CABG by using a novel flow analyzer that can calculate bypass graft resistance (Ra), resistance of the peripheral bed to which graft connects (Rp), the inertia of blood flow through the graft (L) and vascular wall compliance (C). Methods: We performed off-pump CABG surgery on fifteen pigs assigned to the following groups (n = 5 each): normal CABG, competitive flow grafts and constrictive grafts. Results: The wave pattern of 3 groups showed a clearly different form. In normal CABG and competitive flow group, we accepted a statistical difference in Rp and flow. In normal CABG and constrictive grafts. We accepted a statistical difference in Ra and flow. Conclusion: We can know the relationship between CABG and coronary circulation by this device in detail. This device will be useful for evaluating graft performance during CABG. © 2012 The Editorial Committee of Annals of Thoracic and Cardiovascular Surgery. All rights reserved.

Cite

CITATION STYLE

APA

Nakata, K., Orime, Y., Akiyama, K., Koba, H., Sankai, Y., & Shiono, M. (2012). Novel device accurately measures graft resistance and compliance to ensure quality of coronary artery bypass. Annals of Thoracic and Cardiovascular Surgery, 18(5), 438–443. https://doi.org/10.5761/atcs.oa.11.01858

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free