Abstract
For perfusion-based functional magnetic resonance imaging, the previously introduced flow-sensitive alternating inversion recovery (FAIR) technique is combined with single-shot RARE (rapid acquisition with relaxation enhancement) and GRASE (gradient and spin echo) imaging sequences. The advantages of these sequences compared to commonly used echo-planar imaging (EPI) are an increased signal-to-noise ratio and the absence of distortions and artifacts due to magnetic field inhomogeneities. RARE- and GRASE-FAIR are applied to functional brain mapping studies in humans during visual stimulation. Results demonstrate that the presented techniques allow for perfusion maps with higher spatial resolution compared to EPI-FAIR. Relative regional cerebral blood flow change in the occipital cortex during visual stimulation was measured to be 41 ± 4% (n = 5). The comparison of FAIR data obtained with RARE and GRASE techniques shows that RARE yields images with the higher signal-to-noise ratio. However, the GRASE technique features a shorter acquisition time and less RF power deposition and is thus better suited for multi-slice acquisitions.
Author supplied keywords
Cite
CITATION STYLE
Crelier, G. R., Hoge, R. D., Munger, P., & Pike, G. B. (1999). Perfusion-based functional magnetic resonance imaging with single-shot RARE and GRASE acquisitions. Magnetic Resonance in Medicine, 41(1), 132–136. https://doi.org/10.1002/(SICI)1522-2594(199901)41:1<132::AID-MRM18>3.0.CO;2-5