Accuracy of the enlite 6-day glucose sensor with Guardian and Veo calibration algorithms

89Citations
Citations of this article
88Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Objective: This study investigates the accuracy of a newly developed, next-generation subcutaneous glucose sensor, evaluated for 6-day use. Research Design and Methods: Seventy-nine subjects (53 men, 26 women) with type 1 diabetes and 18 subjects (14 men, four women) with type 2 diabetes completed a three-center, prospective, sensor accuracy study. The mean age for the group was 42.2±15.0 years (mean±SD), ranging from 18 to 71 years, with a mean glycosylated hemoglobin level of 7.6±1.5%, ranging from 5.5% to 14%. Subjects wore Enlite ™ sensors (Medtronic Diabetes, Northridge, CA) in the abdominal and buttocks region for two separate 7-day periods and calibrated with a home-use blood glucose meter. Subjects participated in an in-clinic testing day where frequent sampled plasma glucose samples were acquired every 15 min for 10 h. Sensor data was retrospectively processed with Guardian® REAL-Time (Medtronic) and Paradigm® Veo™ (Medtronic) calibration routines, and accuracy metrics were calculated for each algorithm and sensor location. Physiological time lag for each measurement site was calculated. Results: Based on 6,404 plasma-sensor glucose paired points, the Enlite sensor with Veo calibration algorithm produced a mean absolute relative difference of 13.86% with 97.3% of points within the A+B zones of the Clarke error grid. Threshold-only alarms detected 90.1% of hypoglycemia and 90% of hyperglycemia. Mean time lag measured at the abdominal region was 7.94±6.48 min compared with 11.70±6.71 min (P<0.0001) at the buttocks area. Conclusions: The Enlite sensor accurately measures glucose when compared with gold standard laboratory measurements over its 6-day use. Sensors placed in the buttocks region exhibited greater time lags than sensors placed in the abdomen. © 2012, Mary Ann Liebert, Inc.

Cite

CITATION STYLE

APA

Keenan, D. B., Mastrototaro, J. J., Zisser, H., Cooper, K. A., Raghavendhar, G., Lee, S. W., … Shah, R. V. (2012). Accuracy of the enlite 6-day glucose sensor with Guardian and Veo calibration algorithms. Diabetes Technology and Therapeutics, 14(3), 225–231. https://doi.org/10.1089/dia.2011.0199

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free