Abstract
A n 83-year-old man with a history of coronary artery disease and 3-vessel coronary artery bypass grafting underwent implantation of a pacemaker (Medtronic EnPulse E2DR01) at an outside facility because he experienced recurrent syncope and an abnormal response to tilt table testing. Active fixation leads were implanted in the right atrium (Medtronic 5076-52) and right ventricle (Medtronic 5076-58). At the time of implantation, all lead parameters were within normal limits, and the right ventricular lead was positioned in the apex. Twelve days after the implantation, the patient developed intermittent left chest muscle twitching. He presented to an outside hospital's emergency department, where chest radiographs revealed that the right ventricular lead was outside the cardiac silhouette (Figure 1). A trans-thoracic echocardiogram excluded a pericardial effusion. He was subsequently transferred to our institution, where inter-rogation revealed visible evidence of left pectoralis major muscle contraction when pacing the right ventricle at 2.5 V and 0.5 ms. Left chest muscle twitching resolved with reprogramming to AAI mode at a rate of 50 bpm. Chest CT showed the right ventricle electrode tip position to be in the anterior chest wall and confirmed the absence of a pericardial effusion (Figure 2). The patient was taken to the operating room. The right femoral vein and artery were exposed to provide rapid access in the event that cardiopul-monary bypass would be required. Continuous transesopha-geal echocardiography was used to monitor the pericardial space. The ventricular lead was disconnected from the pace-maker generator. Under fluoroscopic guidance, the surgeon retracted the screw into the lead. Gentle, steady traction resulted in lead extraction in its entirety without complica-tions or the development of pericardial fluid. A new ventric-ular lead was not implanted at the time. The patient was discharged home and was doing well at his 1-month follow-up examination. This case illustrates the potential for delayed pacemaker lead perforation after an uneventful implantation of an active fixation lead. 1 Pacing electrode stimulation of the chest wall musculature and absence of a pericardial effusion are unusual features of this presentation.
Cite
CITATION STYLE
Greenberg, S., Lawton, J., & Chen, J. (2005). Right Ventricular Lead Perforation Presenting as Left Chest Wall Muscle Stimulation. Circulation, 111(25). https://doi.org/10.1161/circulationaha.104.494732
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.