Sensing probe for 3-6 GHz microwave imaging systems

9Citations
Citations of this article
7Readers
Mendeley users who have this article in their library.

Abstract

A sensing probe for medical imaging diagnostic systems is presented. The proposed probe consists of a double-ridged waveguide (DRWG) filled with dielectric ceramic and an aperture with a tapered ridge. The DRWG is used to suppress higher-order mode radiations in the rectangular waveguide. The proposed probe is designed for broadband sensing in an immersion liquid with a complex dielectric constant. The performance of the fabricated probe successfully covers the frequency band of 3-6 GHz. The probe suppresses higher-order modes and provides symmetric and similar radiation patterns within the operating frequency range. © 2014 The Institution of Engineering and Technology.

Cite

CITATION STYLE

APA

Lee, K. J., Kim, J. Y., Son, S. H., Lee, J., & Jeon, S. (2014). Sensing probe for 3-6 GHz microwave imaging systems. Electronics Letters, 50(15), 1049–1050. https://doi.org/10.1049/el.2014.1923