Abstract
A method for calibrating iodine-125 seeds in terms of exposure has been established. The standard free-air ionization chamber, used for measuring soft x rays, was chosen for the measurements. Arrays of four to six seeds were used to enhance the ionization-current-to-background-current ratio. Seeds from an array were measured individually in a re-entrant chamber. The quotient of the exposure rate for the array by the sum of the ionization currents in the re-entrant chamber is the calibration factor for the re-entrant chamber. Calibration factors were established for three types of iodine-125 seeds. The overall uncertainty for the seed exposure calibrations is less than 6%.
Author supplied keywords
Cite
CITATION STYLE
Loftus, T. P. (2001). Exposure standardization of iodine-125 seeds used for brachytherapy. Journal of Research of the National Institute of Standards and Technology, 106(5), 857–865. https://doi.org/10.6028/jres.106.042