Abstract
X-ray polarization-contrast images resulting from X-ray linear dichroism (XLD) in the hard X-ray region have been successfully recorded for the first time. The apparatus used consisted of an X-ray polarizer, double X-ray phase retarders and a high-spatial-resolution X-ray charge-coupled device (CCD) detector. The sample used was a hexagonal close packed (h.c.p.) cobalt single-crystal foil of thickness about 12 μm. The experiment was performed at X-ray energies of 23 and 29 eV above the cobalt K edge (7709 eV), at which the maximum linear dichroisms (∼3%) were observed, with their signs reversed, in the XLD spectrum measured with quadruple X-ray phase retarders. The contrasts in the images at the two X-ray energies were reversed as a result of the XLD in the sample. Furthermore, the values of the contrast in the images arising from the linear dichroism (∼3%) were in good agreement with those yielded by the XLD spectrum.
Author supplied keywords
Cite
CITATION STYLE
Sato, K., Okitsu, K., Ueji, Y., Matsushita, T., & Amemiya, Y. (2000). Demonstration of X-ray linear dichroism imaging with hard X-rays. Journal of Synchrotron Radiation, 7(6), 368–373. https://doi.org/10.1107/S0909049500009985