Is computer-assisted surgery in total knee arthroplasty as accurate as it can be?

0Citations
Citations of this article
12Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Three hundred and eighty computer-assisted total knee arthroplasty cases were reviewed for accuracy of mechanical alignment. The 331 patients in the first set, designated Group A, showed a consistent valgus error of 1° from neutral alignment. It was hypothesized that a manual 1° varus correction during femur resection would yield a significantly greater level of accuracy in the second set of 49 patients, designated Group B. A mechanical alignment of ± 3° was achieved in 91% of the uncalibrated Group A patients, which was significantly lower (p = 0.035) than the rate of 98% achieved in the calibrated Group B. Further statistical analysis of the data showed the time expenditure was not significantly changed once a new target value was recalibrated. By quantifying mean errors of measures at an early timeframe, enhanced accuracy in CAS can be achieved.